Impact of the Bayesian penalized likelihood algorithm (Q.Clear®) in comparison with the OSEM reconstruction on low contrast PET hypoxic images.
Ontology highlight
ABSTRACT: PURPOSE:To determine the impact of the Bayesian penalized likelihood (BPL) reconstruction algorithm in comparison to OSEM on hypoxia PET/CT images of NSCLC using 18F-MIZO and 18F-FAZA. MATERIALS AND METHODS:Images of low-contrasted (SBR = 3) micro-spheres of Jaszczak phantom were acquired. Twenty patients with lung neoplasia were included. Each patient benefitted from 18F-MISO and/or 18F-FAZA PET/CT exams, reconstructed with OSEM and BPL. Lesion was considered as hypoxic if the lesion SUVmax > 1.4. A blind evaluation of lesion detectability and image quality was performed on a set of 78 randomized BPL and OSEM images by 10 nuclear physicians. SUVmax, SUVmean, and hypoxic volumes using 3 thresholding approaches were measured and compared for each reconstruction. RESULTS:The phantom and pati