Simulation and Experimental Study on Doubled-Input Capacitively Coupled Contactless Conductivity Detection of Capillary Electrophoresis.


ABSTRACT: In this contribution, we optimize the structure of double-input capacitively coupled contactless conductivity detector (DIC4D) that proposed before by our group and successfully applied it in the capillary electrophoresis of inorganic ion analysis. Furthermore, we present the detail theoretical analysis and simulation to exploring the working mechanism of DIC4D. Compared with C4D, under identical experimental conditions and by using the same current-to-voltage converter, both the theoretical and experimental results suggest that the effectiveness and feasibility of DIC4D. The improved DIC4D diminished the baseline drift effects in C4D, provides lower noise, higher sensitivity and notably stable baseline. The LODs of DIC