ABSTRACT: Objective
The aims of this study were to investigate the feasibility of using a universal abdominal acquisition protocol on a photon-counting detector computed tomography (PCD-CT) system and to compare its performance to that of single-energy (SE) and dual-energy (DE) CT using energy-integrating detectors (EIDs).Methods
Iodine inserts of various concentrations and sizes were embedded into different sizes of adult abdominal phantoms. Phantoms were scanned on a research PCD-CT and a clinical EID-CT with SE and DE modes. Virtual monoenergetic images (VMIs) were generated from PCD-CT and DE mode of EID-CT. For each image type and phantom size, contrast-to-noise ratio (CNR) was measured for each iodine insert and the area under the receiver operating characteristic curve (AUC) f