Optimization of a chest computed tomography protocol for detecting pure ground glass opacity nodules: A feasibility study with a computer-assisted detection system and a lung cancer screening phantom.
Ontology highlight
ABSTRACT: OBJECTIVE:This study aimed to optimize computed tomography (CT) parameters for detecting ground glass opacity nodules (GGNs) using a computer-assisted detection (CAD) system and a lung cancer screening phantom. METHODS:A lung cancer screening phantom containing 15 artificial GGNs (-630 Hounsfield unit [HU], 2-10 mm) in the left lung was examined with a CT scanner. Three tube voltages of 80, 100, and 120 kVp were used in combination with five tube currents of 25, 50, 100, 200, and 400 mA; additionally, three slice thicknesses of 0.625, 1.25, and 2.5 mm and four reconstruction algorithms of adaptive statistical iterative reconstruction (ASIR-V) of 30, 60, and 90% were used. For each protocol, accuracy of the CAD system was evaluated for nine target GGNs of 6, 8, or 10 mm in size. The cut-off