ABSTRACT: Background
Accurate diagnosis of anemia by community workers using a point-of-care device is a challenge. The objective of the study was to establish the diagnostic accuracy of point-of-care devices for detecting anemia in community settings.Methods
It was diagnostic accuracy study with cross-sectional design on adult patients attending the outpatient department of rural/ urban health centres of Medical colleges from India. The index tests were HemoCue, TrueHb, Massimo's device and spectroscopic device, compared against autoanalyzer (gold standard). Accuracy was expressed by sensitivity, specificity, likelihood ratios, predictive values, area under the curve (AUC) and levels of agreement. For the diagnostic accuracy component, 1407 participants were recruited with a minimum