ABSTRACT: Background
Here, we report on a head-to-head comparison of the fully-automated Elecsys® Anti-SARS-CoV-2 immunoassay with the EDITM enzyme linked immunosorbent assays (ELISA) for the detection of SARS-CoV-2 antibodies in human plasma.Methods
SARS-CoV-2 antibodies were measured with the Elecsys® assay and the EDITM ELISAs (IgM and IgG) in 64 SARS-CoV-2 RT-PCR confirmed COVID-19 patients with serial blood samples (n = 104) collected at different time points from symptom onset. Blood samples from 200 healthy blood donors and 256 intensive care unit (ICU) patients collected before the COVID-19 outbreak were also used.Results
In COVID-19 patients, the percentage of positive results rose with time from symptom onset, peaking to positivity rates after 15-22 days of 100% for the Elecsys® assay, of 94% for the EDITM IgM-ELISA and of 100% for the EDITM IgG ELISA. In the 104 blood samples, the agreement between positive/negative classifications of the Elecsys® assay and the EDITM ELISAs (IgM or IgG) was 90%. The false positivity rates in the healthy blood donors and the ICU patients were < 1% for the Elecsys® assay and < 3% for the EDITM ELISAs.Conclusions
Our results indicate a high sensitivity and specificity for the Elecsys® assay and an acceptable agreement with the EDITM ELISAs.
SUBMITTER: Egger M
PROVIDER: S-EPMC7261064 | biostudies-literature | 2020 Oct
REPOSITORIES: biostudies-literature

Clinica chimica acta; international journal of clinical chemistry 20200530
<h4>Background</h4>Here, we report on a head-to-head comparison of the fully-automated Elecsys® Anti-SARS-CoV-2 immunoassay with the EDI<sup>TM</sup> enzyme linked immunosorbent assays (ELISA) for the detection of SARS-CoV-2 antibodies in human plasma.<h4>Methods</h4>SARS-CoV-2 antibodies were measured with the Elecsys® assay and the EDI<sup>TM</sup> ELISAs (IgM and IgG) in 64 SARS-CoV-2 RT-PCR confirmed COVID-19 patients with serial blood samples (n = 104) collected at different time points fro ...[more]