ABSTRACT: Background
Here, we report on a head-to-head comparison of the fully-automated Elecsys® Anti-SARS-CoV-2 immunoassay with the EDITM enzyme linked immunosorbent assays (ELISA) for the detection of SARS-CoV-2 antibodies in human plasma.Methods
SARS-CoV-2 antibodies were measured with the Elecsys® assay and the EDITM ELISAs (IgM and IgG) in 64 SARS-CoV-2 RT-PCR confirmed COVID-19 patients with serial blood samples (n = 104) collected at different time points from symptom onset. Blood samples from 200 healthy blood donors and 256 intensive care unit (ICU) patients collected before the COVID-19 outbreak were also used.Results
In COVID-19 patients, the percentage of positive results rose with time from symptom onset, peaking to positivity rates after