Peripheral cannulae selection for veno-arterial extracorporeal life support: a paradox.


ABSTRACT: Explosive penetration of veno-arterial extracorporeal life support in everyday practice has drawn awareness to complications of peripheral cannulation, resulting in recommendations to use smaller size cannulae. However, using smaller cannulae may limit the effectiveness of extracorporeal support and thereby the specific needs of the patient. Selection of proper size cannulae may therefore pose a dilemma, especially since pressure-flow characteristics at different hematocrits are lacking. This study evaluates the precision of cannula pressure drop prediction with increase of fluid viscosity from water flow-pressure charts by M-number, dynamic similarity law, and via fitted parabolic equation. Thirteen commercially available peripheral cannulae were used in this in vitro study. Pressure drop