Intra-aneurysmal pressure changes during stent-assisted coiling.


ABSTRACT: We aimed to examine aneurysm hemodynamics with intra-saccular pressure measurement, and compare the effects of coiling, stenting and stent-assisted coiling in proximal segments of intracranial circulation. A cohort of 45 patients underwent elective endovascular coil embolization (with or without stent) for intracranial aneurysm at our department. Arterial pressure transducer was used for all measurements. It was attached to proximal end of the microcatheter. Measurements were taken in the parent artery before and after embolization, at the aneurysm dome before embolization, after stent implantation, and after embolization. Stent-assisted coiling was performed with 4 different stents: LVIS and LVIS Jr (Microvention, Tustin, CA, USA), Leo (Balt, Montmorency, France), Barrel VRD (Medtronic/ C