Versatile compact heater design for in situ nano-tomography by transmission X-ray microscopy.
Ontology highlight
ABSTRACT: A versatile, compact heater designed at National Synchrotron Light Source-II for in situ X-ray nano-imaging in a full-field transmission X-ray microscope is presented. Heater design for nano-imaging is challenging, combining tight spatial constraints with stringent design requirements for the temperature range and stability. Finite-element modeling and analytical calculations were used to determine the heater design parameters. Performance tests demonstrated reliable and stable performance, including maintaining the exterior casing close to room temperature while the heater is operating at above 1100°C, a homogenous heating zone and small temperature fluctuations. Two scientific experiments are presented to demonstrate the heater capabilities: (i) in situ 3D nano-tomography including a stu
SUBMITTER: Antonelli S
PROVIDER: S-EPMC7285687 | biostudies-literature | 2020 May
REPOSITORIES: biostudies-literature
ACCESS DATA