Comparing image quality of single- and dual-energy computed tomography of the brain.


ABSTRACT:

Purpose

Weighted average dual-energy computed tomography (DE-CT) reconstructions are considered a proxy of standard CT images of the brain, recommended for routine clinical use and used as a reference standard in DE-CT research. However, their image quality has not been assessed, which was the aim of our study.

Methods

Images from 81 consecutive patients who underwent both non-contrast single-energy (SE)-CT and DE-CT of the brain on the same scanner were retrospectively evaluated. Attenuation values (HU) and SD of grey matter/white matter (GM/WM) pairs, along with SD in the posterior fossa and subcalvarial region were measured. Four readers evaluated image noise, GM/WM contrast, posterior fossa and subcalvarial artefacts, as well as overall image quality.

Results

Weig