Unknown

Dataset Information

Development of an In Situ Analyzer Based on Sequential Injection Analysis and Liquid Waveguide Capillary Flow Cell for the Determination of Dissolved Reactive Phosphorus in Natural Waters.


ABSTRACT: This study presents an innovative technique for the in situ analysis of aquatic biochemical elements detected through wet chemical processes. A new compact in situ phosphate analyzer based on sequential injection analysis, liquid waveguide capillary flow cell and spectrophotometry was developed, and a safe and modular electronics-chemical separation mechanical structure was designed. The sequential injection system of this analyzer was optimized, and the major functions of this analyzer were studied and estimated. With a 10 cm liquid waveguide capillary flow cell and a 6.3 min time cost of detection, the analyzer reaches a detection limit of 1.4 μg·L-1 (≈14.7 nM, [PO43-]) and a consumption of 23 μL at most for each reagent. This analyzer was operated in situ and online during two scientifi

SUBMITTER: Yang Z 

PROVIDER: S-EPMC7288038 | biostudies-literature | 2020 May

REPOSITORIES: biostudies-literature