Reproducibility of Ablated Volume Measurement Is Higher with Contrast-Enhanced Ultrasound than with B-Mode Ultrasound after Benign Thyroid Nodule Radiofrequency Ablation-A Preliminary Study.


ABSTRACT: The reproducibility of contrast-enhanced ultrasound (CEUS) and standard B-mode ultrasound in the assessment of radiofrequency-ablated volume of benign thyroid nodules was compared. A preliminary study was conducted on consecutive patients who underwent radiofrequency ablation (RFA) of benign thyroid nodules between 2014 and 2016, with available CEUS and B-mode post-ablation checks. CEUS and B-mode images were retrospectively evaluated by two radiologists to assess inter- and intra-observer agreement in the assessment of ablated volume (Bland-Altman test). For CEUS, the mean inter-observer difference (95% limits of agreement) was 0.219 mL (-0.372-0.809 mL); for B-mode, the mean difference was 0.880 mL (-1.655-3.414 mL). Reproducibility was significantly higher for CEUS (85%) than for B-mode