Design and ex vivo characterization of narrow implants with custom piezo-activated osteotomy for patients with substantial bone loss.


ABSTRACT:

Objective

Bone augmentation delays implant placement and increases risks due to additional surgeries. Implant systems compatible with reduced alveolar bone volume are required. To design, manufacture, and test a non-cylindrical dental implant system using piezotomes and custom-designed matching titanium mini-implants to address the needs of patients with missing teeth and narrow jawbone.

Materials and methods

Tapered mini-implants with a rectangular cross-section (4.6 mm × 2.1 mm) were machined with dimensions that could accommodate narrow alveolar ridges. The performance of the implants were tested in both static and fatigue cycle 30° compression tests. Tapered, rectangular cutting tools that matched the overall trapezoidal morphology of the implant were also designed. Thes