Continuous LVAD monitoring reveals high suction rates in clinically stable outpatients.


ABSTRACT: Suction of the left ventricle can lead to potentially life-threatening events in left ventricular assist device (LVAD) patients. With the resolution of currently available clinical LVAD monitoring healthcare professionals are unable to evaluate patients' suction occurrences in detail. This study investigates occurrences and durations of suction events and their associations with tachycardia in stable outpatients. Continuous high-resolution LVAD data from HVAD patients were analyzed in the early outpatient period for 15 days. A validated suction detection from LVAD signals was used. Suction events were evaluated as suction rates, bursts of consecutive suction beats, and clusters of suction beats. The occurrence of tachycardia was analyzed before, during, and after suction clusters. Furtherm

SUBMITTER: Gross C 

PROVIDER: S-EPMC7318142 | biostudies-literature | 2020 Jul

REPOSITORIES: biostudies-literature