Ontology highlight
ABSTRACT: Objective
A computational model that accounts for heterogeneous tissue properties was used to compare multiple independent current control (MICC), multi-stim set (MSS), and concurrent activation (co-activation) current steering technologies utilized in deep brain stimulation (DBS) on volume of tissue activated (VTA) and power consumption.Methods
A computational model was implemented in Sim4Life v4.0 with the multimodal image-based detailed anatomical (MIDA) model, which accounts for heterogeneous tissue properties. A segmented DBS lead placed in the subthalamic nucleus (STN). Three milliamperes of current (with a 90 μs pseudo-biphasic waveform) was distributed between two electrodes with various current splits. The laterality, directional accuracy, volume, and shape of the
SUBMITTER: Zhang S
PROVIDER: S-EPMC7318189 | biostudies-literature | 2020 Jun
REPOSITORIES: biostudies-literature