Unknown

Dataset Information

Evaluation of the RealStar® SARS-CoV-2 RT-PCR kit RUO performances and limit of detection.


ABSTRACT:

Background

The use of efficient, reliable and sensitive PCR assays is a cornerstone in the race to contain the SARS-CoV-2 pandemic. In this work we performed an independent evaluation of the RealStar® SARS-CoV-2 RT-PCR Kit Researh Use Only (Altona) for SARS-CoV-2 detection.

Methods

A comparative limit of detection (LoD) assessment was performed between RealStar® SARS-CoV-2 and the currently WHO recommended RT-PCR (WHO-PCR) workflow using a quantified clinical sample. Assessment of the RealStar® SARS-CoV-2 assay was also performed using 83 primary clinical samples in comparison with the WHO-PCR.

Results

The RealStar® SARS-CoV-2 demonstrated a slightly higher sensitivity than the WHO recommended assay with a limit of detection at 625 copies/mL instead of 1250 copies/mL

SUBMITTER: Visseaux B 

PROVIDER: S-EPMC7323686 | biostudies-literature | 2020 Aug

REPOSITORIES: biostudies-literature