The potential of an optical surface tracking system in non-coplanar single isocenter treatments of multiple brain metastases.
Ontology highlight
ABSTRACT: To evaluate the accuracy of a commercial optical surface tracking (OST) system and to demonstrate how it can be implemented to monitor patient positioning during non-coplanar single isocenter stereotactic treatments of brain metastases. A 3-camera OST system was used (Catalyst HD™, C-RAD) on a TruebeamSTx with a 6DoF couch. The setup accuracy and agreement between the OST system, and CBCT and kV-MV imaging at couch angles 0° and 270°, respectively, were examined. Film measurements at 3 depths in the Rando-Alderson phantom were performed using a single isocenter non-coplanar VMAT plan containing 4 brain lesions. Setup of the phantom was performed with CBCT at couch 0° and subsequently monitored by OST at other couch angles. Setup data for 7 volunteers were collected to evaluate the accuracy