ABSTRACT:

Purpose

To develop and evaluate an improved velocity-selective (VS) labeling pulse for myocardial arterial spin labeling (ASL) perfusion imaging that addresses two limitations of current pulses: (1) spurious labeling of moving myocardium and (2) low labeling efficiency.

Methods

The proposed myocardial VSASL labeling pulse is designed using a Fourier Transform based Velocity-Selective labeling pulse train. The pulse utilizes bipolar velocity-encoding gradients, a 9-tap velocity-encoding envelope, and double-refocusing pulses with Malcolm Levitt phase cycling. Amplitudes of the velocity-encoding envelope were optimized to minimize the labeling of myocardial velocities during stable diastole (±2-3 cm/s) and maximize the labeling of coronary velocities (10-130 cm/s during rest/s