Large scale validation of a new non-invasive and non-contact bilirubinometer in neonates with risk factors.


ABSTRACT: The study was aimed to evaluate the performance of a newly developed non-invasive and non-contact bilirubin measurement device (AJO-Neo) as an alternative to the conventional invasive biochemical method of total serum bilirubin (TSB) estimation in preterm and term neonates suffering from hyperbilirubinemia associated with risk factors, and/or undergoing phototherapy. The safety and efficacy of the device were assessed in 1968 neonates with gestational ages ranging from 28 to 41 weeks and suffering from incidences of hyperbilirubinemia. Linear regression analysis showed a good correlation between AJO-Neo and the conventional method of TSB (Pearson's coefficient, r = 0.79). The small bias (0.27 mg/dL) and limits of agreements (- 3.44 to 3.99 mg/dL) were within the range of clinical acceptanc

SUBMITTER: Halder A 

PROVIDER: S-EPMC7341797 | biostudies-literature | 2020 Jul

REPOSITORIES: biostudies-literature