Identification of malignant early repolarization pattern by late QRS activity in high-resolution magnetocardiography.


ABSTRACT:

Background

The early repolarization pattern (ERP) in electrocardiography (ECG) has been considered as a risk for ventricular fibrillation (VF), but effective methods for identification of malignant ERP are still required. We investigated whether high spatiotemporal resolution 64-channel magnetocardiography (MCG) would enable distinction between benign and malignant ERPs.

Methods

Among all 2,636 subjects who received MCG in our facility, we identified 116 subjects (43 ± 18 years old, 54% male) with inferior and/or lateral ERP in ECG and without structural heart disease, including 13 survivors of VF (ERP-VF(+)) and 103 with no history of VF (ERP-VF(-)). We measured the following MCG parameters in a time-domain waveform of relative current magnitude: (a) QRS duration (MCG-QRSD), (b) root-mean-square of the last 40 ms (MCG-RMS40), and (c) low amplitude (<10% of maximal) signal duration (MCG-LAS).

Results

Compared to ERP-VF(-), ERP-VF(+) subjects presented a significantly longer MCG-QRS (108 ± 24 vs. 91 ± 23 ms, p = .02) and lower MCG-RMS40 (0.10 ± 0.08 vs. 0.25 ± 0.20, p = .01) but no difference in MCG-LAS (38 ± 22 vs. 29 ± 23 ms, p = .17). MCG-QRSD and MCG-RMS40 showed significantly larger area under the ROC curve compared to J-peak amplitude in ECG (0.72 and 0.71 vs. 0.50; p = .04 and 0.03). The sensitivity, specificity, and odds ratio for identifying VF(+) based on MCG-QRSD ≥ 100 ms and MCG-RMS40 ≤ 0.24 were 69%, 74%, and 6.33 (95% CI, 1.80-22.3), and 92%, 48%, and 10.9 (95% CI, 1.37-86.8), respectively.

Conclusion

Magnetocardiography is an effective tool to distinguish malignant and benign ERPs.

SUBMITTER: Iwakami N 

PROVIDER: S-EPMC7358799 | biostudies-literature | 2020 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Identification of malignant early repolarization pattern by late QRS activity in high-resolution magnetocardiography.

Iwakami Naotsugu N   Aiba Takeshi T   Kamakura Shiro S   Takaki Hiroshi H   Furukawa Toshiaki A TA   Sato Tosiya T   Sun Wenxu W   Shishido Toshiaki T   Nishimura Kunihiro K   Yamada-Inoue Yuko Y   Nagase Satoshi S   Shimizu Wataru W   Yasuda Satoshi S   Sugimachi Masaru M   Kusano Kengo K  

Annals of noninvasive electrocardiology : the official journal of the International Society for Holter and Noninvasive Electrocardiology, Inc 20200119 4


<h4>Background</h4>The early repolarization pattern (ERP) in electrocardiography (ECG) has been considered as a risk for ventricular fibrillation (VF), but effective methods for identification of malignant ERP are still required. We investigated whether high spatiotemporal resolution 64-channel magnetocardiography (MCG) would enable distinction between benign and malignant ERPs.<h4>Methods</h4>Among all 2,636 subjects who received MCG in our facility, we identified 116 subjects (43 ± 18 years ol  ...[more]