Can QRS morphology be used to differentiate between true septal vs. apparently septal lead placement? An analysis of ECG of real mid-septal, apparent mid-septal, and apical pacing.
Ontology highlight
ABSTRACT: The location of the pacemaker lead is based on the shape of the lead on fluoroscopy only, typically in the left and right anterior oblique positions. However, these fluoroscopy criteria are insufficient and many leads apparently considered to be in septum are in fact anchored in anterior wall. Periprocedural ECG could determine the correct lead location. The aim of the current analysis is to characterize ECG criteria associated with a correct position of the right ventricular (RV) lead in the mid-septum. Patients with indications for a pacemaker had the RV lead implanted in the apex (Group A) or mid-septum using the standard fluoroscopic criteria. The exact position of the RV lead was verified using computed tomography. Based on the findings, the mid-septal group was divided into two subgr