Comparison of electrolyte and glucose levels measured by a blood gas analyzer and an automated biochemistry analyzer among hospitalized patients.


ABSTRACT:

Background

Blood gas analyzers are capable of delivering results on electrolytes and metabolites within a few minutes and facilitate clinical decision-making. However, whether the results can be used interchangeably with values measured by chemistry analyzers remains controversial. Blood gas analyzers are capable of delivering results on electrolytes and metabolites within a few minutes and facilitate clinical decision-making. However, whether the results can be used interchangeably with values measured by chemistry analyzers remains controversial.

Methods

In total, arterial and matched venous blood samples were collected from 200 hospitalized patients. Arterial blood samples were evaluated using a RAPIDPOINT 500 to test electrolyte and glucose levels, then the samples were

SUBMITTER: Yi HC 