Unknown

Dataset Information

The Next Generation of IR Spectroscopy: EC-QCL-Based Mid-IR Transmission Spectroscopy of Proteins with Balanced Detection.


ABSTRACT: We report a mid-IR transmission setup for the analysis of the protein amide I and amide II band in aqueous solutions that achieves a limit of detection as low as 0.0025 mg mL-1 (outperforming our previous results and other state-of-the-art mid-IR-based techniques by almost an order of magnitude). This large improvement is made possible by combining the latest-generation external cavity-quantum cascade laser (EC-QCL) operated at room temperature with an optimized double-beam optical setup that adjusts the path length (26 μm) to ensure robust sample handling. For minimizing the noise introduced by the high-intensity laser light source, a thermoelectrically cooled mercury cadmium telluride balanced detection module was employed. In this way, noise levels better by a factor of up to

SUBMITTER: Akhgar CK 

PROVIDER: S-EPMC7376528 | biostudies-literature | 2020 Jul

REPOSITORIES: biostudies-literature