A Systematic Approach to Improve Scatter Sensitivity of a Flow Cytometer for Detection of Extracellular Vesicles.


ABSTRACT: Extracellular vesicles (EVs) are commonly studied by flow cytometry. Due to their small size and low refractive index, the scatter intensity of most EVs is below the detection limit of common flow cytometers. Here, we aim to improve forward scatter (FSC) and side scatter (SSC) sensitivity of a common flow cytometer to detect single 100 nm EVs. The effects of the optical and fluidics configuration on scatter sensitivity of a FACSCanto (Becton Dickinson) were evaluated by the separation index (SI) and robust coefficient of variation (rCV) of polystyrene beads (BioCytex). Improvement is defined as increased SI and/or reduced rCV. Changing the obscuration bar improved the rCV 1.9-fold for FSC. A 10-fold increase in laser power improved the SI 19-fold for FSC and 4.4-fold for SSC, whereas the r