Sensitivity assessment of droplet digital PCR for SARS-CoV-2 detection.


ABSTRACT: Reverse transcription‑quantitative polymerase chain reaction (RT‑qPCR) is the gold standard method for the diagnosis of COVID‑19 infection. Due to pre‑analytical and technical limitations, samples with low viral load are often misdiagnosed as false‑negative samples. Therefore, it is important to evaluate other strategies able to overcome the limits of RT‑qPCR. Blinded swab samples from two individuals diagnosed positive and negative for COVID‑19 were analyzed by droplet digital PCR (ddPCR) and RT‑qPCR in order to assess the sensitivity of both methods. Intercalation chemistries and a World Health Organization (WHO)/Center for Disease Control and Prevention (CDC)‑approved probe for the SARS‑CoV‑2 N gene were used. SYBR‑Green RT‑qPCR is not able to diagnose as positive samples with low viral