Magic angle effect on adiabatic T1ρ imaging of the Achilles tendon using 3D ultrashort echo time cones trajectory.


ABSTRACT: The protons in collagen-rich musculoskeletal (MSK) tissues such as the Achilles tendon are subject to strong dipolar interactions which are modulated by the term (3cos2 θ-1) where θ is the angle between the fiber orientation and the static magnetic field B0 . The purpose of this study was to investigate the magic angle effect in three-dimensional ultrashort echo time Cones Adiabatic T1ρ (3D UTE Cones-AdiabT1ρ ) imaging of the Achilles tendon using a clinical 3 T scanner. The magic angle effect was investigated by Cones-AdiabT1ρ imaging of five cadaveric human Achilles tendon samples at five angular orientations ranging from 0° to 90° relative to the B0 field. Conventional Cones continuous wave T1ρ (Cones-CW-T