Validation of the Use of Automated and Manual Quantitative Analysis of Corneal Nerve Plexus Following Refractive Surgery.
Ontology highlight
ABSTRACT: Following refractive surgery, the cornea is denervated and re-innervated, hence a reproducible tool to objectively quantify this change is warranted. This study aimed to determine the repeatability and reproducibility of corneal nerve quantification between automated (ACCMetrics) and manual software (CCMetrics) following refractive surgery. A total of 1007 in vivo confocal microscopy images from 20 post-small incision lenticule extraction (SMILE) or post-laser-assisted in situ keratomileusis (LASIK) patients were evaluated by two independent observers using CCMetrics for corneal nerve fibre density (CNFD), corneal nerve branch density (CNBD), and corneal nerve fibre length (CNFL). Intra-observer and inter-observer reproducibility and repeatability, as well as agreement and correlation betw