ABSTRACT: Aims
To validate the diagnostic accuracy of the Augurix SARS-CoV-2 IgM/IgG rapid immunoassay diagnostic test (RDT) for COVID-19.Methods
In this unmatched 1:1 case-control study, blood samples from 46 real-time RT-PCR-confirmed SARS-CoV-2 hospitalized cases and 45 healthy donors (negative controls) were studied. Diagnostic accuracy of the IgG RDT was assessed against both an in-house recombinant spike-expressing immunofluorescence assay (rIFA), as an established reference method (primary endpoint), and the Euroimmun SARS-CoV-2 IgG enzyme-linked immunosorbent assays (ELISA) (secondary endpoint).Results
COVID-19 patients were more likely to be male (61% vs 20%; P = .0001) and older (median 66 vs 47 years old; P < .001) than controls. Whole blood IgG-RDT results showed