ABSTRACT: Purpose
Several studies show that media opacity reduces vessel density (VD) measured by image processing algorithms of optical coherence tomography angiography (OCTA). However, different models of OCTA designed their own algorithms and computational methods, which may have different effects of opacity on VD. This study is aimed to investigate the impact of a simulated model of media opacity on quantitative measurement of two OCTA devices.Methods
A spectral-domain based OCTA (Cirrus 5000; Carl Zeiss Meditec) and a swept-source based OCTA (Triton DRI-OCT, Topcon Inc.) were used to image retinal microvasculature at the macula using 3 × 3 mm scanning protocol from 22 eyes of 22 healthy subjects. Media opacity was simulated with neutral-density filters (optical density (OD)