Liquid Cathode Glow Discharge as an Excitation Source for the Analysis of Complex Water Samples with Atomic Emission Spectrometry.


ABSTRACT: A liquid cathode glow discharge (LCGD) was developed as a low-power and miniaturized excitation source of atomic emission spectrometry (AES) for the determination of K, Na, Ca, and Mg in water samples from rivers and lakes. The discharge stability and parameter influencing the analytical performance of LCGD-AES were systematically examined. Moreover, the measurement results of water samples using LCGD-AES were verified by ion chromatography (IC). The results showed that the optimized operating parameters are a 660 V discharge voltage, pH = 1.0 HNO3 as the supporting electrolyte, and a 4.0 mL min-1 solution flow rate. High concentrations of some metals may interfere with the detection of Ca and Mg. Low-molecular-weight organic substances do not have a remarkable enhancement on signal intens