Intradevice Repeatability and Interdevice Agreement of Ocular Biometric Measurements: A Comparison of Two Swept-Source Anterior Segment OCT Devices.


ABSTRACT:

Purpose

To assess the repeatability and agreement of ocular biometric parameters measured using the Tomey CASIA SS-1000 and Heidelberg ANTERION anterior segment optical coherence tomography (AS-OCT) devices.

Methods

Both eyes of subjects 18 years of age or older were scanned three times with the CASIA and ANTERION under standardized dark lighting. One AS-OCT image along the horizontal (temporal-nasal) meridian was analyzed per eye and per scan. Pupillary diameter (PD) was within 15% for all pairwise comparisons. Anterior chamber depth, lens vault, anterior chamber width, angle opening distance, trabecular iris space area, and scleral spur angle (SSA500) were measured using manufacturer-provided image analysis software. Intraclass correlation (ICC), Wilcoxon signed-rank, and