Miniaturized Piezo Force Sensor for a Medical Catheter and Implantable Device.
Ontology highlight
ABSTRACT: Real-time monitoring of intrabody pressures can benefit from the use of miniaturized force sensors during surgical interventions or for the recovery period thereafter. Herein, we present a force sensor made of poly(vinylidene fluoride)-co-trifluoroethylene (P(VDF-TrFE)) with a simple fabrication process that has been integrated into the tip of a medical catheter for intraluminal pressure monitoring, as well as into an implantable device with a power consumption of 180 μW obtained by the near-field communication (NFC) interface to monitor the arterial pulse at the subcutaneous level (≤1 cm). The pressure range supported by the sensor is below 40 kPa, with a signal responsivity of 0.63 μV/Pa and a mean lifetime expectancy of 400 000 loading cycles inside physiological conditions (12 k
SUBMITTER: Gil B
PROVIDER: S-EPMC7450887 | biostudies-literature | 2020 Aug
REPOSITORIES: biostudies-literature
ACCESS DATA