Scaffold underexpansion and late lumen loss after bioresorbable scaffold implantation: Insights from ABSORB JAPAN trial.


ABSTRACT:

Background

Device underexpansion is associated with late adverse outcomes after bioresorbable vascular scaffold (BVS) implantation. This study, representing official IVUS results of the ABSORB Japan trial, aimed to characterize IVUS findings, focusing specifically on acute device expansion, and to investigate its impact on late lumen loss (LLL) with Absorb-BVS compared with cobalt-chromium everolimus-eluting stents (CoCr-EES).

Methods

ABSORB Japan enrolled 148 patients (2:1 randomization) in the IVUS cohort. Serial IVUS was prescheduled at post-procedure and 3 years. Acute device expansion was evaluated with respect to the degree and uniformity of the implanted device.

Results

Overall, Absorb-BVS showed smaller and more nonuniform device expansion at post-procedure, c