Application of high field magnetic resonance microimaging in polymer gel dosimetry.


ABSTRACT:

Purpose

The purpose of this work was to examine the suitability of VIPARnd polymer gel-9.4 T magnetic resonance microimaging system for high spatial resolution dose distribution measurements.

Methods

The VIPARnd samples (3 cm in outside diameter and 12 cm in height) were exposed to ionizing radiation by using a linear accelerator (Varian TrueBeam, USA; 6 MV x-ray beam). In the calibration stage, nine gel dosimeter vials were irradiated in a water phantom homogenously to the doses from 1.5 to 30 Gy in order to obtain R2-dose relation. In the verification stage, two gel dosimeter vials were irradiated in the half beam penumbra area of 10 × 10 cm radiation field using the amount of monitor units appropriate to deliver 20 Gy at the field center. The gels we