Evaluation and Field Calibration of a Low-Cost Ozone Monitor at a Regulatory Urban Monitoring Station.
Ontology highlight
ABSTRACT: The performance of a low cost ozone monitor (Aeroqual Series 500 portable gas monitors using a metal oxide sensor for ozone; model OZL) was assessed under field conditions. Ten ozone monitors were calibrated under clean-air laboratory conditions and controlled ozone concentrations of 5 to 100 ppb. Good linearity and response were obtained relative to a research-grade ozone monitor. One monitor was co-located at a regulatory air quality monitoring station that uses a U.S. federal equivalent method (FEM) ozone analyzer. Raw data from the Aeroqual monitor collected over 4 months (June-October) at a 10-minute time-resolution, showed good agreement (r2=0.83) with the FEM values but with an overestimation of ~12%. Data were averaged to different time resolutions; 1 h time averaged con