Repeatability, reproducibility, agreement, and safety of Tono-Pen tip cover for intraocular measurement using latex and polyethylene wrap.


ABSTRACT:

Purpose

To evaluate repeatability, reproducibility, and agreement of intraocular pressure measurement with Tono-Pen using Ocufilm and polyethylene wrap tip cover in human eyes.

Methods

This is a cross-sectional, experimental study. A gas-sterilized, polyethylene wrap was used as an alternative for Tono-Pen tip cover. For the right eye, 4 measurements using polyethylene wrap tip cover were done by two examiners (A and B) in random order to assess intra-observer repeatability and inter-observer reproducibility. For the left eye, 4 measurements were done by examiner A using both polyethylene wrap tip cover and Ocufilm in random order to assess intra-observer repeatability and agreement. Bland-Altman plot and intraclass correlation coefficient (ICC) were used in all analyses. Co