Unknown

Dataset Information

Numerical model to predict and compare the hypotensive efficacy and safety of minimally invasive glaucoma surgery devices.


ABSTRACT:

Purpose

To predict and compare the hypotensive efficacy of three minimally-invasive glaucoma surgery (MIGS) implants through a numerical model.

Methods

Post-implant hypotensive efficacy was evaluated by using a numerical model and a computational fluid dynamics simulation. Three different devices were compared: the XEN 45 stent (tube diameter, 45 μm), the XEN 63 stent (63 μm) and the PreserFlo microshunt (70 μm). The influence of the filtration bleb pressure (Bp) and tube diameter, length, and position within the anterior chamber (AC) on intraocular pressure (IOP) were evaluated.

Results

Using baseline IOPs of 25, 30 and 50 mmHg, respectively, the corresponding computed post-implant IOPs for each device were as follows: XEN 45: 17 mmHg (29% decrease), 19 mmHg (45%) an

SUBMITTER: Kudsieh B 

PROVIDER: S-EPMC7523982 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature