There is a shortage of chemical reagents for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) diagnosis and a surge of SARS-CoV-2 cases, especially in limited-resource settings. Therefore, the combination of an optimal assay kit is necessary.Methods
We compared the ability to screen SARS-CoV-2 among three primer-probe sets in two different master mixes, Invitrogen™ SuperScript™ III One-Step RT-PCR and LightCycler Multiplex RNA Virus Master.Results
The assay with TIB-Molbiol, IDT, and Phu Sa sets for LightCycler Multiplex RNA Virus Master or Invitrogen™ SuperScript™ III One-Step RT-PCR showed positive results from a single reaction of triplicate in the three days of 4.8 copies per reaction. R squared and amplification efficiency were 0.97 and ranged from 107 to 108%, respectively.Conclusions
Our findings indicated that TIB-Molbiol, IDT, and Phu Sa primer-probe sets could be beneficial for the laboratory screening of SARS-CoV-2 by RT-qPCR assay of E gene. There is a need to consider the combination of these reagent sets as a new strategy to increase the testing capacity of screening programs for COVID-19.
SUBMITTER: Cuong HQ
PROVIDER: S-EPMC7528141 | biostudies-literature | 2020
REPOSITORIES: biostudies-literature
Cuong Hoang Quoc HQ Hai Nguyen Duc ND Linh Hoang Thuy HT Anh Nguyen Hoang NH Hieu Nguyen Trung NT Thang Cao Minh CM Thao Nguyen Thi Thanh NTT Lan Phan Trong PT
BioMed research international 20200925
<h4>Background</h4>There is a shortage of chemical reagents for severe acute respiratory syndrome coronavirus 2 (<i>SARS-CoV-2</i>) diagnosis and a surge of <i>SARS-CoV-2</i> cases, especially in limited-resource settings. Therefore, the combination of an optimal assay kit is necessary.<h4>Methods</h4>We compared the ability to screen <i>SARS-CoV-2</i> among three primer-probe sets in two different master mixes, Invitrogen™ SuperScript™ III One-Step RT-PCR and LightCycler Multiplex RNA Virus Mas ...[more]