Validation of the Polar V800 heart rate monitor and comparison of artifact correction methods among adults with hypertension.
Ontology highlight
ABSTRACT: Heart rate variability (HRV) measurements via ambulatory monitors have become common. We examined the validity of recording R-R intervals using the Polar V800™ compared to 12-lead electrocardiograms (ECG) among middle-aged (44.7±10.1years); overweight to obese (29.8±4.3 kg.m-2) adults (n = 25) with hypertension (132.3±12.2/ 84.3±10.2 mmHg). After resting for 5-min in the supine position, R-R intervals were simultaneously recorded using the Polar V800™ and the 12-lead ECG. Artifacts present in uncorrected (UN) R-R intervals were corrected with the Kubios HRV Premium (ver. 3.2.) automatic (AC) and threshold-based (TBC) correction, and manual correction (MC) methods. Intra-class correlation coefficients (ICC), Bland-Altman limits of agreement (LoA), and effect sizes (ES) were calculated. We d