Unknown

Dataset Information

Establishment of sex-specific reference intervals for automated haematology analyser-delivered research parameters in healthy Korean adults: a retrospective database review.


ABSTRACT:

Objectives

Automated haematology analysers measure various parameters of relevance to clinical research along with routine complete blood count (CBC)-related components. We aimed to establish ethnicity-specific and sex-specific reference intervals for 26 research-specific parameters as well as 18 routinely reported components using a large cohort of healthy Korean adults. The necessity of requiring separate sex-specific reference intervals for each parameter was also examined.

Design

A retrospective database review.

Setting

Single tertiary-care hospital of approximately 375 physicians and 530 nurses.

Participants

This study included 1383 reference individuals (840 men and 543 women).

Primary and secondary outcome measures

Following the Clinical and Labor

SUBMITTER: Jeon K 

PROVIDER: S-EPMC7545637 | biostudies-literature | 2020 Oct

REPOSITORIES: biostudies-literature