IFCC Working Group Recommendations for Correction of Bias Caused by Noncommutability of a Certified Reference Material Used in the Calibration Hierarchy of an End-User Measurement Procedure.


ABSTRACT: Establishing metrological traceability to an assigned value of a matrix-based certified reference material (CRM) that has been validated to be commutable among available end-user measurement procedures (MPs) is central to producing equivalent results for the measurand in clinical samples (CSs) irrespective of the clinical laboratory MPs used. When a CRM is not commutable with CSs, the bias due to noncommutability will be propagated to the CS results causing incorrect metrological traceability to the CRM and nonequivalent CS results among different MPs. In a commutability assessment, a conclusion that a CRM is commutable or noncommutable for use with a specific MP is made when the difference in bias between the CRM and CSs meets or does not meet a criterion for that specific MP when compare

SUBMITTER: Miller WG 

PROVIDER: S-EPMC7551486 | biostudies-literature | 2020 Jun

REPOSITORIES: biostudies-literature