A Novel Method for Measuring Serum Unbound Bilirubin Levels Using Glucose Oxidase-Peroxidase and Bilirubin-Inducible Fluorescent Protein (UnaG): No Influence of Direct Bilirubin.
Ontology highlight
ABSTRACT: The glucose oxidase-peroxidase (GOD-POD) method used to measure serum unbound bilirubin (UB) suffers from direct bilirubin (DB) interference. Using a bilirubin-inducible fluorescent protein from eel muscle (UnaG), a novel GOD-POD-UnaG method for measuring UB was developed. Newborn sera with an indirect bilirubin/albumin (iDB/A) molar ratio of <0.5 were classified into four groups of DB/total serum bilirubin (TB) ratios (<5%, 5-10%, 10-20%, and ≥20%), and the correlation between the UB levels and iDB/A ratio was examined. Linear regression analysis was performed to compare UB values from both methods with the iDB/A ratio from 38 sera samples with DB/TB ratio <5% and 11 samples with DB/TB ratio ≥5%. The correlation coefficient (r) between UB values and the iDB/A ratio for the GOD-POD method