Unknown

Dataset Information

Spray-Capillary: An Electrospray-Assisted Device for Quantitative Ultralow-Volume Sample Handling.


ABSTRACT: The analysis of low-volume samples provides valuable insight into complex biological systems. However, the proteomic and metabolomic analysis of low-volume samples remains challenging due to the lack of simple, efficient, and reproducible microsampling techniques. We have developed an electrospray-assisted device for quantitative low-volume sample extraction, referred to here as "Spray-Capillary". Stable electrospray was achieved through a chemically etched tip from a long (e.g., 50 cm) capillary with a conductive sheath flow. This electrospray provided the driving force to quantitatively draw low-volume samples into the capillary. We evaluated the precision and accuracy of sample injection volumes using our spray-capillary as the electrospray voltage, capillary ID, and column length were

SUBMITTER: Huang L 

PROVIDER: S-EPMC7558432 | biostudies-literature | 2020 Jan

REPOSITORIES: biostudies-literature