Ontology highlight
ABSTRACT: Objectives
Curing lights cannot be sterilized and should be covered with an infection control barrier. This study evaluated the effect of barriers when applied correctly and incorrectly on the radiant power (mW), irradiance (mW/cm2), emission spectrum (mW/nm), and beam profile from a multi-peak light-curing unit (LCU).Methods
Five plastic barriers (VALO Grand, Ultradent; TIDIShield, TIDI Products; Disposa-Shield, Dentsply Sirona; Cure Sleeve, Kerr; Stretch and Seal, Betty Crocker) and one latex-based barrier (Curelastic, Steri-Shield) were tested. The radiant power (mW) and emission spectrum (mW/nm) from one multi-peak LCU (VALO Grand, Ultradent) was measured using an integrating sphere. LCU tip internal diameter (mm) was measured, then the tip area and irradianc
SUBMITTER: Soares CJ
PROVIDER: S-EPMC7569381 | biostudies-literature | 2020 Dec
REPOSITORIES: biostudies-literature