A Simulation Study to Assess the Effect of Analytic Error on Neonatal Glucose Measurements Using the Canadian Pediatric Society Position Statement Action Thresholds.


ABSTRACT:

Background

The Canadian Pediatric Society (CPS) has endorsed an algorithm for the screening and immediate management of babies at risk of neonatal hypoglycemia that provides time-dependent glucose concentration action thresholds. The objective of this study was to evaluate the impact of glucose analytic error (bias and imprecision) on the misclassification of glucose meter results from a neonatal intensive care unit (NICU) using the CPS guidelines.

Methods

A simulation dataset of true glucose values (N = 100 000) was derived by finite mixture model analysis of NICU glucose data (N = 23 749). Bias and imprecision were added to create measured glucose values. The percentages of measured glucose values that were misclassified at CPS action thresholds were determin