Unknown

Dataset Information

The Exchange-Free Technique: A Novel Technique for Enhancing Surpass Flow Diverter Placement.


ABSTRACT:

Background

Following flow diverter placement, approximately 20% of intracranial aneurysms remain as residual aneurysms at 1 year. Device malapposition is a cause of residual aneurysms after flow diversion. We present a new and straightforward technique (exchange-free technique [EFT]) to enhance apposition of the surpass flow diverter (SFD), the only over-the-wire flow diverter currently available.

Materials and methods

We deployed laser-cut mini stents through the inner deployment catheter of the SFD. This maneuver was performed simply by withdrawing the micro-guidewire from its lumen and replacing it with a mini-stent (stents deliverable through microcatheters with inner diameter of 0.0165 inches), without a need to re-cross the deployed SFD or an exchange maneuver. All ane

SUBMITTER: Ocal O 

PROVIDER: S-EPMC7591168 | biostudies-literature | 2020 Jul-Sep

REPOSITORIES: biostudies-literature