ABSTRACT: Purpose
Conventional hemodialysis (HD) treatment has an acceptable removal of small uremic molecules, but so-called "middle molecules" in the range of 0.5-60 kDa are poorly cleared with HD compared to a native kidney, which may contribute to morbidity in the dialysis population. Hemodiafiltration (HDF) has a better removal of middle molecules compared to HD but is technically demanding and requires well-functioning dialysis access. The newly introduced medium cutoff (MCO) filters have been developed to enhance middle molecule clearance in HD-mode. The aim of this study was to compare reduction ratios (RRs) of molecules with different molecular weights (0.06-150 kDa) during dialysis with MCO dialyzer (used in HD-mode) compared to online-hemodiafiltration (ol-HDF) treatment with a co