The comparison of analytical performances of Mindray CL-1000i and Beckman Coulter Access II Troponin I methods in the light of recent guidelines and the quality requirements.


ABSTRACT:

Background

The results of cardiac troponin I (cTnI) methods used in the diagnosis of acute myocardial infarction (AMI) are highly variable. In this study, it was aimed to compare the analytical performance of the Mindray CL-series TnI method with the Beckman Coulter-Access II AccuTnI+3 method.

Methods

Analytical performance and method comparison experiments for cTnI measurement with Mindray CL-1000i and Beckman Coulter-Access II instruments were with the Clinical and Laboratory Standards Institute (CLSI) documents EP15-A3 and EP9-A3. Precision studies were performed with commercially available third-party quality control (QC) materials. Method comparison experiments were performed by using patient samples. Furthermore, the limit of quantification (LoQ), total analytical erro