Quantifying Precision Loss in Targeted Metabolomics Based on Mass Spectrometry and Nonmatching Internal Standards.


ABSTRACT: In mass spectrometry, reliable quantification requires correction for variations in ionization efficiency between samples. The preferred method is the addition of a stable isotope-labeled internal standard (SIL-IS). In targeted metabolomics, a dedicated SIL-IS for each metabolite of interest may not always be realized due to high cost or limited availability. We recently completed the analysis of more than 70 biomarkers, each with a matching SIL-IS, across four mass spectrometry-based platforms (one GC-MS/MS and three LC-MS/MS). Using data from calibrator and quality control samples added to 60 96-well trays (analytical runs), we calculated analytical precision (CV) retrospectively. The use of integrated peak areas for all metabolites and internal standards allowed us to calculate precisio