Comparing the robustness of different skin flash approaches using wide tangents, manual flash VMAT, and simulated organ motion robust optimization VMAT in breast and nodal radiotherapy.


ABSTRACT: Compare the robustness of wide tangents (WT) and volumetric modulated arc therapy (VMAT) using different skin flash approaches in breast and nodal radiotherapy. Ten patients treated with WT using 2-cm flash were replanned with VMAT using no flash (NF), manual 2-cm flash (MF), and robust optimization (RO). Plan robustness was assessed for target coverage and organs at risk (OAR) by recalculating on 5 deformed CT scans (SOM1-5), daily cone beam (CBCT), and by shifting the isocenter 5 mm. VMAT NF gave poor coverage of CTVp with its smallest change of -3.2% for V38Gy on CBCT. VMAT RO plans showed the least variations in target coverage loss compared to WT and VMAT MF which dropped as anatomical swelling increased. CTVp D0.5cc decreased on CBCT and increased most for VMAT